SHOPRIDER SERVICE MANUAL 2007
1-800-743-0772
Stop the wheelchair and reverse down the gradient. While on the gradient release the joystick and
ensure the wheelchair comes to rest and the brakes are applied without the front wheels liftin g off the
ground.
NO POWER TEST
No Power to Joystick Controller
Check Circuit Breaker an d Reset
If OK…
Check both Fuses inline on B attery Cables
If OK…
Check for 24v Voltage at Female connector of Main Wiring Harness:
Using a volt meter set on VDC, put one probe in #1 & one probe in #2. Your reading
should be approximately 24v. Next put one probe in #3 & one probe in #4. Your reading
should be ap proximately 24v.
If all the above is OK, Replace VSI Controller.
1
2
3
4